Description

Sensor Head Spot Type, Laser Class 2
Model LK-H152
Mounting mode Diffuse reflective / Specular reflective*1
Reference distance Diffuse reflective: 150 mm Specular reflective: 147.5 mm
Measuring range Diffuse reflective: ±40 mm Specular reflective: -39.5 mm to +24.4 mm*2
Light source Type Red semiconductor laser
Wavelength 655nm
Laser class (JIS C6802) Class 2
Output 0.95mW
Spot diameter (at standard distance) o120 µm
Linearity ±0.02% of F.S. (F.S.= 80 mm)*3
Repeatability 0.25 µm*4
Sampling cycle 2.55/5/10/20/50/100/200/500/1000 µs (9 steps selectable)
Temperature characteristics 0.01% of F.S./°C (F.S.= 80 mm)
Environmental resistance Enclosure rating IP67
Ambient light Incandescent lamp or fluorescent lamp: 5,000 lux max.
Ambient temperature 0 to +50 °C*5
Relative humidity 35 to 85 % RH (No condensation)
Vibration resistance 10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z directions
Material Aluminium die-cast
Weight Approx. 300 g
*1 Use one of the following dark filters when measuring a transparent or mirror surface object: LK-H082/LK-H087: LK-F3, LK-H152/LK-H157: LK-F2 *2 Measurement range when the sampling cycle is 20 µs or more. *3 This value is obtained when the KEYENCE standard target (White diffuse workpiece or workpiece with a metal mirror surface only for the LK-H008/LK-H008W) is measured in the normal measurement mode. *4 This value is obtained when the KEYENCE standard target (White diffuse workpiece or workpiece with a metal mirror surface only for the LK-H008/LK-H008W) is measured at the reference distance with the number of averaging measurements set to 16384. The value in parentheses is a typical e x ample of a measurement with the number of averaging measurements set to 65536 and the sampling cycle to 200 µs. *5 When the ambient temperature rises to 40 °C or more, mount this on a metal plate before use.
''