Description

Controller
Model LS-9501*1
Type Controller: NPN output type
No. of connectable sensor heads 2
Head compatibility Yes
Display Minimum display unit 0.01 µm
Display range ±99,999.99 µm to ±9,999.9 mm
LED display POWER ON indicator, ERROR indicator
Input terminal block Encoder input NPN/PNP open-collector output, voltage output (5 V/12 V/24 V), line-driver output
Synchronous 1, 2 input Non-voltage input
Auto-zero 1, 2 input
Reset 1, 2 input
Storage trigger input
Storage enable input
Storage data clear input
Statistics 1, 2 input
Statistics clear 1, 2 input
Program selection input Non-voltage input x 4 inputs
Output terminal Analogue voltage output ±10 V x 2 outputs, output impedance 100 ?
Analogue current output 4 to 20 mA x 2 outputs, compatible load max. 350 ?
Universal output NPN open-collector output x 10 outputs Measured value and tolerance judgment output, status output allocatable
Status 1, 2 output NPN open-collector output
Total judgment output
Memory FULL output
Strobe 1, 2 output
Error output NPN open-collector output (N.C.)
Interface Ethernet 1000BASE-T/100BASE-TX
USB USB 2.0 HI-SPEED supported (USB 1.1 Full-SPEED compatible)
RS-232C Measured value output, control I/O, setting change, baud rate can be selected up to 115,200 bps
Display and settings panel LS-D1000 Max. four heads connectable
Rating Power voltage 24 VDC ±10 %, including Ripple (P-P)
Current consumption When LS-HA100 not used: 1.0 A max. when 1 head connected, 1.4 A max. when 2 heads connected When LS-HA100 in use: 2.0 A max. when 3 heads connected, 2.3 A max. when 4 heads connected*2
Environmental resistance Ambient temperature When LS-HA100 not used: 0 to +50 °C When LS-HA100 in use: 0 to +45 °C
Relative humidity 20 to 85 % RH (No condensation)
Weight Approx. 1,500 g
*1 NPN open-collector output rating: 50 mA max. (40 V max.), residual voltage of 1 V max. Non-voltage input rating: ON voltage of 1 V max., OFF current of 0.6 mA max. *2 Add the current consumption values for all units when connecting the display settings panel and expansion units. When the LS-9006D or LS-9030D is connected, it counts as two heads.
''